Asepti Autozyme Pr
Multi-enzymatic medical instrument detergent class IIb enzymatic prior deactivator biofilm remover
ARTG R 156033
![]() |
![]() |
![]() |
![]() |
![]() |
|
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |
Features
- Asepti Autozyme PR standardizes reprocessing by providing prion deactivation as an integrated component of all instrument cleaning, alleviating the inherent problems associated with risk classification of medical instrument loads.
- Asepti Autozyme PR offers superior deactivation of human M1000 prions to WHO protocols and removes Biofilm (validated as per ISO 15883-5 Annex F).
- Asepti Autozyme Pr is suitable for all automated and ultrasonic washers offering validated efficacy for deactivation of human M1000* prions with universal material compatibility.
- * Human M1000 Prion Strain — this prion strain has proven to be significantly more resistant to chemical and thermal deactivation than the scrapie-derived 263K prion strain used to validate World Health Organisation recommended prion deactivation cycles.
- Asepti Autozyme Pr is fully compatible with MIS and robotic instruments.
| Product Code | Product Description | Pack Size | Quote |
|---|---|---|---|
Safety Information
For further information and safety instructions see product label and Safety Data Sheet.
ALWAYS READ THE LABEL. FOLLOW DIRECTIONS OF USE.
Directions
Instrument Cleaning (Routine Instrument Reprocessing):
- Dosage 3 – 5 mL/Litre. Temp: 55°C – 65°C. Recommended contact time: Not less than 5 minutes.
- Note contact time to be adjusted in line with in-situ performance qualification. Minimum contact time: Not less than 2 minutes.
Combined Cleaning and greater than LOG2 reduction of Prion titre*:
- Dosage: 4 – 10 mL/Litre. Temp: 60°C – 65°C. Contact time: not less than 10 minutes.
- Rinse well, follow by autoclaving when possible to achieve combined > 4log reduction in prion titre.
Combined Cleaning and Deactivation of Prions*:
- Dosage: 4 – 10 mL/Litre. Temp: 60° – 65°C. Contact time: not less than 30 minutes.
- Rinse well, follow by autoclaving or chemical disinfection.
Combined Cleaning and Complete Deactivation of Prions*:
- Dosage: 4 – 10 mL/Litre. Temp: 60°C – 65°C. Contact time: not less than 30 minutes.
- Rinse well. Autoclave: autoclave for at least 20 minutes at 121°C to achieve complete M1000 prion deactivation.
* WHO recommends combination of at least two prion inactivating methods. Avoiding heating detergent solutions above 70°C as enzymes are deactivated above this temperature.
Applications
- Dental
- Infection Control
- Healthcare Facilities
- Central Sterile Departments
Downloads
To access a copy of the eBook or product brochure, please complete and submit a copy of the Downloads form below.