Soluscope P
Endoscope and Tee Probe Disinfectant
ARTG R 118778
![]() |
![]() |
![]() |
Features
- Peracetic acid-based formula
- Proven antimicrobial activity in minutes in the Soluscope AER
- Compliant with EN ISO 15883-4
- Compatible with all types of flexible endoscopes
- Specific connection to avoid confusion with the additive and the cleaner for the washer-disinfector (AER)
| Product Code | Product Description | Pack Size | Quote |
|---|---|---|---|
Active Against
- Bacteria: EN 13727 + A2, EN 14561 Contact Time: 5 minutes at 40°C
- Mycobacteria: EN 14348, EN 14563 Contact Time: 5 minutes at 40°C
- Yeast/Molds: EN 13624, EN 14562 Contact Time: 5 minutes at 40°C
- Viruses: EN 14476 + A1 Contact Time: 5 minutes at 40°C
- Bacterial spores: D-Value Contact Time: 5 minutes at 40°C
Composition
- Peracetic Acid (5%)
- Hydrogen Peroxide
- Acetic Acid
Characteristics
- Clear, colourless solution
- Density: between 1.10 and 1.13
- pH (pure product) between 1.00 and 4.20
- Stability: 14 months (closed), 2 months (opened)
Safety Information
Do not mix with any other products. Further information, safety instructions and health warnings, see product label and Safety Data Sheet.
ALWAYS READ THE LABEL. FOLLOW DIRECTIONS OF USE.
Directions
- Concentrated solution for professional use exclusively in Soluscope Automated Endoscope Reprocessors: Sprint and Sprint PT
- Always read the label and follow the directions for use. Do not mix with any other products.
- Don the required PPE.
- Screw the plunger, equipped with the red cap, onto the bottle.
- Make sure that the vent hole is not blocked.
- Put the bottle in place.
- Store between 5°C and 25°C.